This page contains information about changes/issues for RHESSI detector 4 after the 2014 anneal. Detector updates from 2012-mid 2014 can be found here: Detector 4 anneal 2012
Contents |
Notes
After the 2014 anneal, detector 4 is UNSEGMENTED.
- On May 21 starting at 19:24 UTC we raised up the Slow threshold for the Front segment.
15-141-19:24:10 start idib_chg_thrshld (4, FRONT, SLOW, 0x68) --> Was 0x60 15-141-19:24:22 /IDPUDUMPTABL TABLE=DIBTBL4 15-141-19:24:22 /IDPUTABLE4 OFFSET=FRONTSLOWDAC 15-141-19:24:31 /IDPLOAD VALUE=0x68
D4 front slow LLD threshold history
post 2014 anneal | 0x50 |
May 21 2015 19:24 UTC | 0x70 (was 0x60) |
D4 front fast LLD threshold history
post 2014 anneal | 0x30 |
2014-Dec-8 after 22:05 UTC | 0x50 (was 0x30) |
D4 rear slow LLD threshold history
post 2014 anneal | 0x30 |
D4 rear fast LLD threshold history
post 2014 anneal | 0x45 |
D4 HV history
post 2014 anneal | 4510 V |