From RHESSI Wiki
This page contains information about changes/issues for RHESSI detector 5 after the 2014 anneal. Detector updates from 2012-mid 2014 can be found here: Detector 5 anneal 2012
Notes
- After the 2014 anneal, detector 5 is UNSEGMENTED.
- A tweak shortly after the anneal was completed:
G5 front slow to 0x80 (was 0x70) on 2014 Aug. 14 after 03:40 UTC
- In late December 2014 some weak signal of segmentation were seen on G5. On January 5, 2015 at 14:57 UTC we set new values for the front fast and front slow thresholds on this detector. We need to keep an eye on its behaviour. Detector 5 is now weakly SEGMENTED.
G5 front slow to 0x0C (was 0x80) on 2015 Jan. 5 after 14:57 UTC
G5 front fast to 0x2E (was 0x60) on 2015 Jan. 5 after 14:57 UTC
- On 2014 Jan 8, the rear fast threshold was raised since fast counts were eating up a lot of livetime.
15-008-17:44:15 /IDPULOAD VALUE=0x50 ;(was 0x45)
D5 front slow LLD threshold history
post 2014 anneal | 0x70
|
2014-Aug-14 after 03:40 UTC | 0x80 (was 0x70)
|
2015-Jan-5 after 14:57 UTC | 0x0C (was 0x80)
|
D5 front fast LLD threshold history
post 2014 anneal | 0x30
|
2014-Dec-8 after 20:28 UTC | 0x60 (was 0x30)
|
2015-Jan-5 after 14:57 UTC | 0x2E (was 0x60)
|
D5 rear slow LLD threshold history
D5 rear fast LLD threshold history
post 2014 anneal | 0x45
|
2014-Jan-08 17:44:15 | 0x50 (was 0x45)
|
D5 HV history